Model 450 Gaussmeter
Features
Measures DC or AC magnetic fields
Features Max Hold, Autorange, Alarm, Relative and Filter modes
Resolution to 5 ¾ digits (1 part in ±300,000)
DC accuracy is ±0.10% of reading
High-visibility vacuum-fluorescent display is easily read at a distance
Analog voltage outputs, RS232C serial port, and
IEEE-488 port
Includes axial or transverse probe
Description:
The Model 450 is a high-resolution, laboratory gaussmeter, capable of measuring magnetic fields from 0.001 Gauss to 300k Gauss (measurement range is dependent on probe used). This unit is designed for demanding applications, such as material analysis and field mapping, where low-noise, high-resolution magnetic measurements are necessary. Measurements of AC and DC magnetic fields are shown in Gauss or Tesla on the two-line vacuum fluorescent display. The Model 450 features RS232C and IEEE-488 ports, allowing the instrument to be controlled via a remote computer or automation equipment. A hall-effect gaussmeter probe is included with the Model 450, and may be selected from an extensive line of probe constructions.
Price: $3069.00
price subject to change without notice.
General Specifications | ||
Measurement Ranges | 300mG
/ 3 G / 30 G / 300 G / 3 kG / 30 kG / 300 kG (range is dependent on probe type) |
|
Resolution | 0.001mG (when used with filter on and appropriate probe) | |
Accuracy, DC (displayed) | ± 0.10% of reading; ± 0.005% of range | |
Accuracy, AC (displayed) | ± 2% of reading, rms (50 Hz – 60 Hz) | |
Frequency Range | 10 Hz – 400 Hz | |
Update Rate | 5 readings / sec (display); up to 18 readings / sec on ports | |
Display Resolution | ± 5 3/4 digits (max.) | |
Measurement Modes | DC, RMS, Peak | |
Measurement Features | Auto Range, Max Hold, Relative Mode, Filter | |
Measurement Units | Gauss or Tesla (with u, m, and k multipliers) | |
Probes | Features | Linearity
correction, auto probe zero, temperature correction |
Types | Axial, transverse, flexible transverse | |
Compatibility | Standard and custom probes available | |
Connector | 15-pin D-style | |
Front Panel Display | Vacuum-Flourescent, high-visibiliy blue, 2 lines x 20 characters | |
Communication Ports | RS232C
(300 / 1200 / 9600 Baud) IEEE-488 (complies with IEEE-488.2) |
|
Alarms | Settings | High/low set point, inside/outside, audible |
Actuators | Display annunciator, beeper | |
Temperature Range | 15 – 35° C (operating temp; 5 – 40° with reduced accuracy) | |
Input Power | 100, 120, 220, 240 VAC 50/60Hz (select voltage when ordering) |
Analog Outputs | ||
Analog
Outputs 1 (monitor output ) |
Configuration | Real-time analog output voltage |
Output Voltage | ± 3V = ± FS on selected range | |
Accuracy | Probe dependent | |
Frequency Response | DC to 400 Hz | |
Analog Output
2 (corrected output) |
Configuration | Output voltage from DAC |
Output Voltage | ± 3V (Model 450-10: ± 10V) | |
Accuracy | ± 0.1% of FS | |
Update Rate | 5 updates / sec | |
Connectors | 2 x BNC jacks |
Physical | |
Net Weight | 6.6 lbs (3 kg) |
Size (W x H x D) | 8.5″ x 3.5″ x 12.5″ (217mm x 90mm x 317mm; fits in half-rack space) |
Package Contents | Gaussmeter,
manual, zero-gauss chamber, line cord, probe (select from axial, transverse, or flexible transverse models) |
Software | Labview driver (available for download) |